Oxidation
of Cysteine to Cystine
Important Functional Groups in this experiment
R-NH2 amine
R-CO2H carboxylic
acid
R-CONH2 amide
R-S-H thiol
R-S-R sulfide
R-S-S-R disulfide

[Ox] = H2O2,
hydrogen peroxide
Proteins are natural polymers made up of
amino acids linked by peptide bonds.
Naturally Occurring Amino Acid

With the exception of the amino acid glycine, where R = H, all amino acids are chiral.
If R is a lower priority than -CO2H
and NH2, the absolute configuration is S.

Wade: Table 24-2 p. 1155-6 lists the 20 amino acids commonly
found in proteins.

In proteins, the amino group of 1
amino acid is attached to the carboxylic acid group of another amino acid.

The
amide functional group is formed.
peptide: dipeptide
peptide bond
Proteins are often referred to as
polypeptides.


R= any of the structures found for the
naturally occurring amino acids.

disulfide
bond

separate polypeptide chains
If the thiol
groups of the cysteine units (residue) on two
different polypeptide chains are oxidized, the result is the formation of a
disulfide between the 2 polymer chains.
The 2 polymer chains are now linked,
and the material is said to be a crosslinked polymer.
Crosslinked polymers tend to be stronger
materials than single strand polymers?
If there is a relatively small amount of crosslinking,
the material is elastic (natural rubber, neoprene, etc.). If there is a large amount of crosslinking, the material becomes tough and had strong
mechanical properties.

Hair contains the protein keratin,
which contains a relatively high percentage of cystine.
Decomposition
Some compounds decompose before they melt.
Usually, at the decomposition point, the solid darkens rapidly, and bubbling is
observed. Using a sealed m.p. tube, the
decomposition point of cystine should be observed at
235º C. The method for sealing
m.p. tubes will be demonstrated to small groups.
The IR spectrum will be obtained. One way of obtaining IR spectra of
solids is to use a Nujol Mull.
IR
Spectra of Solids
1.
KBr pellet
2.
Nujol Mull
3.
Solution IR
1.
Grind 5 mg. of sample into a fine powder using a mortar and pestle.
2.
Add 1-2 drops of Nujol (mineral oil).
3.
Grind it again until the mixture is a fine dispersion. (Nujol
Mull) (Looks like thin milk)
4.
Place the mull between the salt plates.
5.
Obtain the IR spectrum.
6. Clean the plates with methylene chloride and hexanes.
Nujol (mineral oil): mixture of high m.w. alkanes
Absorb » in same regions as C-H vibrations of sample.
(See Fig. 19-10 in manual, p. 854)
The Nujol
peaks obscure any peaks that might be due to C-H vibration of sample. Mark
peaks due to Nujol.
If too much Nujol
is used, then the peaks of the compound being analyzed will be hard to
distinguish from even the weakest peaks Nujol.
Amino acids are found in 3 forms.
1. free amino acid (zwitterion)

A. 3100 - 2600 cm-1 (s,
broad) NH3+ stretching vibration + other complex
vibrations with one sharp band between 2222 - 2000 cm-1
B. 1660 - 1590 cm-1 (w) NH3+
asymmetric bending
1550 - 1485 cm-1 (s) NH3+
symmetric bending
C. 1650 - 1590 cm-1 (s)
asymmetric stretch of carboxylate anion
1400 cm-1 (w) symmetric
stretch of carboxylate anion
2. hydrochloride (or other salt)

A. 3333 - 2380 cm-1
superimposed N-H and O-H stretching vibrations
B. 1610 - 1590 cm-1 (w) NH3+
asymmetric bending vibration
1550 - 1481 cm-1 (s) NH3+
symmetric bending vibration
C. 1220 - 1190 cm-1 (s) C-O
stretching in the carboxylic acid
D. 1755 - 1730 cm-1 (s) C=O
stretch for amino acids
3. Sodium (or other cation) salts

A. 3400 - 3200 cm-1 N-H
stretch normally found in amines
B. 1600 - 1590 cm-1 , 1400
cm-1 vibrations due to carboxylate anion
You will be asked to identify whether
the product is in the free amino acid, hydrochloride, or metal salt form.
Three spectra have been posted near
the IR instrument. One is a typical
background spectrum. A
spectrum of Nujol is provided. It shows the absorptions of Nujol so that you won't confuse them for the peaks
associated with your product cystine. Finally, the third spectrum is that of cystine in Nujol. A good spectrum of your product should
look something like it.


When you look at them, notice that the
Nujol peak near 3000 cm-1 is off
scale. Normally no peak should be
cut off like that in a good spectrum.
However, since the actual absorptions of cystine
are all substantially weaker than the Nujol peaks in
that region and since we are not concerned with Nujol,
it is acceptable change the display limits so that the peaks around 1600 cm-1
appear larger. None of those peaks are off scale in the
example.
They are not labeled in the example,
but for the report, some of the peaks must be labeled. This can be partially accomplished by
using the Find option. For cystine, you will want to
set the threshold so that all the peaks between 1300 and 1650 cm-1
will be listed. For the posted
spectrum, a good threshold value would be 40. At that value, the small peak ~ 2100 cm-1
will not be listed and you will need to label it yourself (using the mouse and
keyboard). Also, you will also have
to separately label the broad absorption ~ 2700 cm-1.
Note some of the peaks may not match
up perfectly. The above values are
general positions for amino acids.
In the report, you will match your peaks up as best as possible to determine
the form of the amino acid present.
To get the best results, err on the
side of above 5 mg. cystine when making the Nujol mull.
KBr does not absorb between 4000 and 600
cm-1.
A solid solution of the sample in KBr can be prepared by placing a physical mixture of the
substances under pressure.
The sample will be held in a thin disk
of KBr (pellet).
KBr Pellet Preparation
1. Dry KBr
by storing in an oven. (If H2O present, broad absorption band
between 3300 - 3500 cm-1.)
KBr is currently stored in the ovens in the
prep room.
2. Mix 1 - 2 mg of sample with 100 mg KBr.
3. Grind the mixture (5 min.) into
fine powder in an agate mortar and pestle.
4. Place portion of sample in die.
5. Tighten bolts (30-40 foot pounds on
torque wrench, too much can shear bolts)
6. Leave the die under pressure for 60
sec.
7. Carefully loosen and remove bolts.
A translucent disk (KBr pellet) should remain in the
center of the die.
8. If the pellet is clear enough,
place it in the holder.
9. Obtain IR spectrum.
Reasons for Cloudy Pellets
1. KBr
mixture not ground enough
2. Sample was not dry
3. Sample: KBr
ratio high
4. Pellet too thick
5. Bolts not tightened sufficiently.
6. Sample has low m.p.
The ends of the bolts are polished
surfaces. Avoid scratching.